Maintenance Menu Items
Create Diagnostic Backup
Thermo Scientific
TriPlus RSH User Guide
185
Create Diagnostic Backup
By using the command
Create Diagnostic Backup
, essential information from the current
status of the TriPlus RSH is extracted, and saved in an encrypted form on the SD card.
WARNING
As pointed out in the
Warning Message
the configuration data will be lost.
This means that in the software, Configuration data base will be deleted and the few
parameters such as the X-Axis length and the Signal settings (see
and
)
will be reset to default and re-entered.
All other parameters from the TriPlus RSH system configuration (e.g. teaching points for
modules, and so on) will be deleted. Consequently, must follow all the steps in the
Installation Wizard
again.
Path from Start Screen:
Options | Service | Installation
.
Access level
Extended User
is required.
Table 86.
List Items for Reset Configuration
Item
Description
Factory Setting 1-S
This will set the length of the X-Axis to 850 mm.
Factory Setting 2-L
This will set the length of the X-Axis to 1205 mm
Factory Setting 3 ITSP-S This will set the basic configuration for the dedicated
application ITSP.
The dedicated Tray Types and the Waste Container for deposing
the used ITSP cartridge are configured besides the standard
parameters as listed below.
Table 87.
System Parameters affected by Reset Configuration
Item
Default Value Settings with Reset Configuration
I/O Interface
TTLIn1 and TTLIn2; negative logic; 20 ms Debounce Time
SWOut1 and SWOut2: PWMOut 36V.
Signals
Input Signals: TTLIn1 and TTLIn2
Output Signals: SWOut1 and SWOut2
Chromatography System GC1: Input1, Output1
Plunger Check
Activated
(Used in Service Installation Wizard Calibrate Tool.)