3
Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
121
Tool Stations
The following stations are actively supported:
•
TR Station
— passive module with single slot for a single tool. See the section
.
•
ATC Station
— active module with three slots for three tools. See the section
.
Upon selecting the list item
Tool Station,
the following parameters are available for
modification. See
. Press
Enter
to edit a parameter.
Figure 84.
Menu Item Tool Stations and Menu List Items for Tool Stations
Adjust Needle Penetration
Allows adjusting the
Needle Penetration Depths
in the
vial. Manually turn the needle as far down in the vial as
required. The cut-out in the Tray Holder frame as well as in
the Rack allows visual control of the needle tip.
Note
The value found is not stored automatically.
This task allows for a check only. If the needle penetration should
be changed, select the Rack Type to enter the newly-established
Needle Penetration value.
New Rack
Allows creating a new Rack
New Well Plate
Allows creating a new Well Plate
Delete Tray Type
Allows deleting the present Tray Type.
Table 50.
Option Pull-up Menu for Temperature Controlled Drawer (Sheet 2 of 2)
Item
Description
Note
TR Station is not supported in combination with the RSH Firmware version
1.5.0 or higher.