Injectors
Upon selecting the list item
Injectors
, the following parameters are available for modification.
See
Figure 79.
Injectors - Selection and Parameters
Press
Enter
to edit a parameter. See
Table 41.
Injector Parameters (Sheet 1 of 2)
Item
Description
Default Needle Penetration
Depth
Defines the
Needle Penetration Depth
in the injector.
It reflects the total length including the length of the fiber.
Max Needle Penetration
Depth
Sets the upper limit of
Needle Penetration Depth
into the
injector.
Min Needle Penetration
Depth
Sets the minimum limit of
Needle Penetration Depth
into
the injector.
WARNING
This parameter is used for
SPME
1
and
SPME 2
Tools only. This parameter is not used for
LS
,
HS,
or
ITEX
Tools. The default value is 18 mm. The adjustable
range is from 10 mm to 50 mm.
Position Tolerance
Checks the difference between the taught X-Y-Axes positions
against the actual position each time the head moves to the
GC Injector. If the difference is greater than the
Position
Tolerance
of 3.0 mm (default value) the TriPlus RSH
immediately stops. This is a safety feature to avoid needle
bending.
WARNING
It is highly recommended to check the teaching
positions of the injectors again when the GC is ready (heated
up to the level as used for routine runs).