Adding Modules
Adding a Generic Detector Interface
484
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
b. Keeping the module flap cover open, place the module in its seat. Note that the
25-pin female connector on the detector seat of the detector housing is not used.
c.
Use a T20 Torxhead screwdriver to tighten the three captive fixing screws without
overtightening.
.
d. Close the module flap cover.
Connecting the Detector Gas Tubings to the Manifolds
This sections provides instruction for connecting the detector gas tubings to the
manifolds of the GDI mechanical module and GDI electrical interface.
To connect the gas tubing block to the manifolds
shows manifolds of the GDI mechanical module and GDI electrical interface.
Figure 514.
Detector Gas Tubings Manifolds (1)
1. Take one, two, or three segments of the 1/16-in.stainless steel tubings (provided)
according to the detector gases required and long enough to properly connect both the
detector gas tubings manifolds.
ATTENTION
To maintain the correct alignment the screws must be tightened in turn.
Tighten each screw only a small amount before moving to the next screw. Repeat until all
are secure.
, and
do not show
the third-party detector for graphic convenience. Suppose that it is installed in the
detector module GDI.
Detector Gas Tubings
Manifolds
Third-party
Detector