9
Adding Modules
Adding a Generic Detector Interface
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
485
2. Connect the detector gas tubings to the manifold of the GDI mechanical module.
See
Figure 515.
Detector Gas Tubings Connections
a.
Connect the detector gas tubings to the numbered inlet ports using the appropriate
Swagelok® 1/16-in. nut and ferrules. Use a 1/4-in. wrench to tighten the fittings.
b. Repeat
until all the required detector gas tubings are connected to the GDI
mechanical module.
c.
Bend and run the detector gas tubings along the top cover.
Third-party
Detector
Detector Gases
Tubings
Detector Gases
Tubings Manifold
ATTENTION
The inlet ports of the manifold on the GDI mechanical module are
numbered
1
,
2
, and
3
respectively. Pay attention to the correct order when you connect
each detector gas tubing to the corresponding outlet ports marked
Gas 1
,
Gas 2
, and
Gas 3
on the GDI electrical interface.
Connect the detector gas tubings for the third-party detector as follows:
• Outlet port
Gas 1
to inlet port
1
• Outlet port
Gas 2
to inlet port
2
• Outlet port
Gas 3
= Air (Wall) to inlet port
3
Note
The bending of the detector gas tubings shown in
is
indicative.