2.4
Solvent and Additive Information
2.4.1
General Compatibility
To protect optimal functionality of the Vanquish system, observe these
recommendations on the use of solvents and additives:
•
The system must be used with reversed-phase (RP) compatible
solvents and additives only.
•
Use only solvents and additives that are compatible with all parts in
the flow path.
TIP
In a Vanquish Core system, normal-phase (NP) compatible solvents
and additives may be used if the VC-pumps and the VC-autosamplers are
modified with the components from the Normal-Phase (NP) kit. Refer to
the
Operating Manuals
for the pumps and autosamplers.
NOTICE
If the system flow path includes a Vanquish charged aerosol detector,
observe the specific solvent and additive information for this detector.
Refer to the
Operating Manual
for the Vanquish charged aerosol
detector.
Summary of Contents for thermoscientific Vanquish UHPLC System
Page 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Page 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Page 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Page 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Page 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Page 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Page 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Page 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Page 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Page 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...