3
Maintenance
Lamp Maintenance
Thermo Scientific
Surveyor UV/Vis Plus Detector Hardware Manual
43
Installing a New Tungsten Lamp
shows the tungsten lamp assembly.
Figure 34.
Tungsten lamp assembly
To install a new tungsten lamp
1. Remove the used tungsten lamp as described in
2. Hold the lamp assembly so that the leads are at the top. Slide the assembly onto the two
alignment pins shown in
. The alignment pins are located on either side of the
detector monochromator aperture.
Figure 35.
Alignment pins for the tungsten lamp
Tungsten bulb
Alignment holes
Alignment pins