Maintenance
Lamp Maintenance
Thermo Scientific
Surveyor UV/Vis Plus Detector Hardware Manual
41
Installing a New Deuterium Lamp
To install a new deuterium lamp
1. Remove the used deuterium lamp as described in
2. Hold the deuterium lamp assembly so that the leads are at the top.
3. Slide the assembly onto the alignment pin (see
).
Figure 32.
Deuterium lamp assembly alignment pin
4. Securely fasten the assembly in place with the two thumbscrews and standoffs.
5. Connect the lamp lead to the upper terminal in the lamp compartment.
6. Replace the lamp housing and secure it with the screw.
7. Close the front doors of the detector.
8. Plug the power cord back into the power receptacle on the back panel of the detector.
Replacing the Tungsten Lamp
To replace the tungsten lamp, follow these steps:
1.
2.
Installing a New Tungsten Lamp
Alignment pin