1
Introduction
Theory of Operation
Thermo Scientific
Surveyor RI Plus Detector Hardware Manual
7
shows a schematic of the optical system. The numbered components shown in
are described in
.
Figure 6.
Optical system components
The condensing lens (item 2) and slit no.1 (item 3) form an image of the tungsten filament
(item 1). This image passes through the collimator lens (item 4) and slit no.2 (item 5), which
project the image of slit no. 2 through the flowcell (item 6) and onto the photodiode (item 9).
Table 4.
Optical system components
Item No.
Description
1
Tungsten lamp
2
Condensing lens
3
Slit No. 1
4
Collimator lens
5
Slit No. 2
6
Flowcell
7
Mirror
8
Null glass
9
Photodiode
1
2
3
4
5
6
7
9
8