Routine and Preventive Maintenance
Maintaining the Probe Heater
Thermo Scientific
Surveyor MSQ Plus Hardware Manual
75
3. Pull the probe heater out of the probe mount, and then disconnect the green connector
and the white connector from the MS detector. See
.
The green connector connects the probe heater to the temperature sensor. The white
connector connects the probe heater to its power source.
4. Store the probe heater in the holder to the left of the source compartment until you are
ready to clean it. See
.
5. To remove the probe heater from the MS detector, use a #1 Phillips head screwdriver to
detach the grounding strap from the probe heater. See
.
Figure 57.
View of power connector and thermocouple connector
Figure 58.
View of probe heater with the operator detaching the grounding strap
White,
Power connector
Green,
Thermocouple connector