5
Operation
Starting operation
5-4
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
5. Do not exceed the maximum level mark “MAX” (1/
). The maximum fill volume of the
water reservoir (5/
) is 3 l.
Figure 5-2.
“MAX” fill level indicator
6. Wipe off excess water from the cover of the water reservoir.
7. Reinstall pre-filter (2/
).
8. Reinstall the bottom shelf or close flap over cut-out for filling.
9. Make sure that the valves of the CO
2
/O
2
/N
2
supply system are open.
10. Turn on the device using the power switch.
11. Adjust set values for temperature and CO
2
/O
2
concentration on the iCan™ touchscreen.
Starting the device:
12. Start the device with the auto-start routine (see
“Activating the auto-start routine“
).
13. The auto-start progress indicator appears on the display, and the automatic start routine is run.
14. The temperature control adjusts the temperature to the selected set value, humidity rises.
15. When temperature and relative humidity are constant, the automatic adjustment of the CO
2
-/O
2
measuring system is performed.
16. The CO
2
-/O
2
control supplies the set amount of CO
2
-/O
2
.
17. If the auto-start routine has been completed, the progress indicator is hidden, and the main menu
is displayed. The device is operational.
1
2
5
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Page 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...