Maintenance
Inspections and checks
To ensure the operational performance and safety of the device, their functions and the components
listed below must be checked at regular intervals.
Daily check:
• Gas reserve of the CO
2
supply system.
• Gas reserve of the O
2
-/N
2
supply system.
Annual inspection:
• Tightness of the glass door seal.
• Permeability of the pressure compensation opening with insert.
• Perform functional check of the control panel and of the incubator’s built-in controller.
• Electrical safety check in accordance with the relevant national regulations (e.g. BGV 3).
Table of Contents
•
•
•
“Preparing Temperature Calibration“
•
“Temperature Calibration Procedure“
•
“Preparing the CO2 calibration“
•
•
•
“Replacing the gas inlet filter“
•
•
Note
Functional check:
If safety devices were removed or disabled for inspections, the device must not be operated before
the safety devices have been reinstalled and checked for their correct function.