Thermo Scientific Star A326 Instruction Sheet Download Page 8

Orion Star A326 Portable pH/RDO/DO Meters

DO Probe Type Selection

The Orion Star A326 meter accepts and automatically recognize Orion RDO optical dissolved oxygen probes 
and Orion polarographic dissolved oxygen probes. If a different probe is used or the probe type needs to be 
verified, perform the following steps.

1.  In the measurement mode, press 

.

2. Press 

  or 

 to highlight 

DO Channel

 and press 

f3 (select)

.

3. Press 

 

or

 

 to highlight 

Mode and Settings

 and press 

f3 (select)

.

4. Press 

 

or

 

 to highlight 

Measure Mode

 and press 

f3 (select)

.

5. Press 

 

or

 

 to highlight 

DO 

(polarographic probe) or 

RDO 

(RDO optical probe) and press 

f3 (select)

.

6. Press 

 to return to the measurement mode.

Viewing the Calibration Log

1.  In the measurement mode, press 

2. Press 

 or 

 to highlight 

View Log

 and press 

f3 (select)

.

3. Press 

 

or

 

 to highlight 

Calibration Log 

and press 

f2 (accept)

.

4. Press 

 

or

 

 to highlight 

pH – Channel 

or

 DO – Channel 

and press 

f2 (select)

.

5. Press 

 

or

 

 to highlight 

pH

,

 RmV

 or

 ORP 

(pH - Channel) or

 DO 

or

 RDO 

(DO- Channel) and press

 f2 

(select)

.

6.  The meter will display a list of calibrations for the selected channel and calibration type. The list shows the 

sequential number of the calibration and the date and time it was saved (

07/01/2011 12:45

).

7.  To view the calibration data, press 

 

or

 

 to highlight a specific calibration and press 

f2 (select)

Press 

f2 (print)

 to print the calibration, press 

f3 (info)

 to view the electrode slope between pH buffer points 

(pH - Channel only) or press 

f1 (back)

 to return to the list of calibrations.

8. Press 

 to return to the measurement mode.

Viewing the Data Log

1.  In the measurement mode, press 

.

2. Press 

 or 

 to highlight 

View Log

 and press 

f3 (select)

.

3. Press 

 

or

 

 

to highlight 

Data Log 

and press

 

(accept)

.

4. Press 

 

or

 

 

to highlight 

pH – Channel 

or

 DO – Channel 

and press 

f2 (select)

.

5.  The meter will display a list of the data points. The list shows the sequential number of the data point and 

the date and time the data point was saved (

07/01/2011 12:45

).

6.  To view the measurement information for an individual data point, press 

 

or

 

 to highlight the data 

point and press 

f2 (select)

. Press 

f2 (print)

 

to print the data point or press 

f1 (back)

 to return to the list of 

data points.

7. Press 

 to return to the measurement mode. 

North America 

166 Cummings Center 
Beverly, MA 01915  USA 
Toll Free: 1-800-225-1480 
Tel: 1-978-232-6000 
[email protected]

Netherlands

Tel: (31) 033-2463887 
[email protected]

India

 

Tel: (91) 22-4175-8800
wai.asia@thermofisher.com 

©2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks 
are the property of Thermo Fisher Scientific Inc. & its subsidiaries. 

Water Analysis Instruments

 

www.thermoscientific.com/water

China

Tel: (86) 21-68654588  
wai.asia@thermofisher.com

Singapore

 

Tel: (65) 6778-6876
wai.asia@thermofisher.com 

Japan

 

Tel: (81) 045-453-9175
wai.asia@thermofisher.com 

Australia

 

Tel: (613) 9757-4300
in Australia (1300) 735-296
InfoWaterAU@thermofisher.com

9 0 0 1 : 2 0 0 8

68X576606  RevA  0811

Summary of Contents for Star A326

Page 1: ...m the meter use a coin or your finger and release the bottom portion of the battery cover d Orientate the batteries as shown in the battery compartment housing and insert batteries e Replace the battery compartment cover and screws Electrodes and Other Connections 1 Prepare the pH electrode RDO optical dissolved oxygen probe or polarographic dissolved oxygen probe and any other applicable electrode...

Page 2: ... used pH Indicates the type of measurement and determines the type of calibration that will be performed ready Specifies the stability of the electrode as stabilizing or ready Shown when the meter is in AUTO READ mode The icon will blink while the reading is stabilizing and stop blinking when the reading is stable and the measurement is locked on the display 7 000 pH Displays the measurement value ...

Page 3: ...e to turn the display backlight on or off or hold down to turn the meter off In the measurement mode press to take a measurement In the setup calibration and other menus press to escape the current menu and return to the measurement mode In the measurement mode press to enter the setup menu In the setup calibration and other menus press to scroll up through a list of options In the continuous meas...

Page 4: ...he numeric entry screen ii Press f2 done to exit the numeric entry screen iii Press f2 accept to accept the entered value 5 Pressf2 next toproceedtothenextbufferandrepeatsteps2through4orpressf3 caldone tosaveand endthecalibration Iffivebuffersareused thecalibrationwillsaveandendoncethefifthvalueisaccepted a If a one point calibration is performed press f2 accept to accept the displayed slope value o...

Page 5: ...o Sat Calibration and 100 0 if using an RDO optical DO probe or 102 3 if using a polarographic DO probe 6 Press f3 cal done to export the data to the calibration log or press f2 print to export the data to the calibration log and a printer or computer The meter will proceed to the measurement mode Set Zero Calibration A zero point calibration is performed in an oxygen free solution A zero point ca...

Page 6: ... taking a measurement to manually export the measurement to the data log if the data log is enabled in the setup menu 1 Rinse the pH electrode RDO optical dissolved oxygen probe or polarographic dissolved oxygen probe and any other electrodes in use with distilled water blot dry with a lint free tissue and place into the sample 2 Start the measurement and wait for it to stabilize a If the meter is...

Page 7: ...t Data t Data Log t Date Time t Language t Key Press Beep t Alarm Beep t Contrast t Auto Shut Off t User ID t Sample ID t Data Log t Calibration Log t Meter Self Test t Factory Reset t About Meter Read Type Selection 1 In the measurement mode press 2 Press or tohighlightpHChannelorDOChannelandpressf3 select 3 Press or to highlight Mode and Settings and press f3 select 4 Press or to highlight Read ...

Page 8: ...n press f3 info to view the electrode slope between pH buffer points pH Channel only or press f1 back to return to the list of calibrations 8 Press to return to the measurement mode Viewing the Data Log 1 In the measurement mode press 2 Press or to highlight View Log and press f3 select 3 Press or to highlight Data Log and press accept 4 Press or to highlight pH Channel or DO Channel and press f2 ...

Reviews: