Thermo Scientific Star A326 Instruction Sheet Download Page 4

Orion Star A326 Portable pH/RDO/DO Meters

4

pH Calibration

One to five pH buffers can be used for calibration. Always use fresh pH buffers and select buffers that bracket 
the sample pH and are one to four pH units apart. Prepare the pH electrode according to the instructions in 
the electrode use guide. Connect the pH electrode and any other electrodes to be used (ATC probe, reference 
electrode) to the meter. Power on the meter and set the measurement mode to pH.

1.  In the measurement mode, press 

f1 (cal)

. Press 

 

or

 

 to highlight 

pH-Channel

 and press 

f2 (select)

.

2.  Rinse the pH electrode and any other electrodes in use with distilled water, blot dry with a lint-free tissue 

and place into the pH buffer.

3.  When the electrode and buffer are ready, press 

f3 (start)

.

4.  Wait for the pH value on the meter to stabilize and stop flashing and perform one of the following actions:

a. Press 

f2 (accept)

 to accept the displayed value.

b. Press 

f3 (edit)

 to access the numeric entry screen and edit the value.

i. Press 

 or 

 to highlight a number, decimal point or negative sign; press 

f3 (enter)

 to select the highlighted item and repeat until the value at the measured temperature 

is shown above the numeric entry screen. 

ii. Press 

f2 (done)

 to exit the numeric entry screen. 

iii. Press 

f2 (accept)

 to accept the entered value.

5. Press 

f2 (next)

 

to proceed to the next buffer and repeat steps 2 through 4 or press 

f3 (cal done)

 to save and 

end the calibration. If five buffers are used, the calibration will save and end once the fifth value is accepted.

a.  If a one point calibration is performed, press 

f2 (accept)

 to accept the displayed slope value or press 

f3 (edit)

 to access the numeric entry screen, enter the slope value and press 

f2 (accept)

6.  The meter will display the calibration summary including the average slope. Press 

f1 (meas)

 to export 

the data to the calibration log or press 

f2 (print)

 

to export the data to the calibration log and a printer or 

computer. The meter will automatically proceed to the measurement mode.

pH Buffer Group Selection

The selected buffer group allows for the automatic recognition of certain pH buffers during a pH calibration. 
The USA buffer group includes pH 1.68, 4.01, 7.00, 10.01 and 12.46 buffers and the DIN buffer group includes 
pH 1.68, 4.01, 6.86, and 9.18 buffers.

1.  In the measurement mode, press 

.

 

2. Press 

 or 

 to highlight 

pH Channel

 and press 

f3 (select)

.

3. Press 

 

or

 

 to highlight 

Mode and Settings 

and press 

f3 (select)

.

4. Press 

 

or

 

 to highlight 

Buffer Group 

and press 

f3 (select)

.

5. Press 

 

or

 

 keys to highlight 

USA

 or

 DIN

 and press 

f3 (select)

6. Press 

 to return to the measurement mode.

Summary of Contents for Star A326

Page 1: ...m the meter use a coin or your finger and release the bottom portion of the battery cover d Orientate the batteries as shown in the battery compartment housing and insert batteries e Replace the battery compartment cover and screws Electrodes and Other Connections 1 Prepare the pH electrode RDO optical dissolved oxygen probe or polarographic dissolved oxygen probe and any other applicable electrode...

Page 2: ... used pH Indicates the type of measurement and determines the type of calibration that will be performed ready Specifies the stability of the electrode as stabilizing or ready Shown when the meter is in AUTO READ mode The icon will blink while the reading is stabilizing and stop blinking when the reading is stable and the measurement is locked on the display 7 000 pH Displays the measurement value ...

Page 3: ...e to turn the display backlight on or off or hold down to turn the meter off In the measurement mode press to take a measurement In the setup calibration and other menus press to escape the current menu and return to the measurement mode In the measurement mode press to enter the setup menu In the setup calibration and other menus press to scroll up through a list of options In the continuous meas...

Page 4: ...he numeric entry screen ii Press f2 done to exit the numeric entry screen iii Press f2 accept to accept the entered value 5 Pressf2 next toproceedtothenextbufferandrepeatsteps2through4orpressf3 caldone tosaveand endthecalibration Iffivebuffersareused thecalibrationwillsaveandendoncethefifthvalueisaccepted a If a one point calibration is performed press f2 accept to accept the displayed slope value o...

Page 5: ...o Sat Calibration and 100 0 if using an RDO optical DO probe or 102 3 if using a polarographic DO probe 6 Press f3 cal done to export the data to the calibration log or press f2 print to export the data to the calibration log and a printer or computer The meter will proceed to the measurement mode Set Zero Calibration A zero point calibration is performed in an oxygen free solution A zero point ca...

Page 6: ... taking a measurement to manually export the measurement to the data log if the data log is enabled in the setup menu 1 Rinse the pH electrode RDO optical dissolved oxygen probe or polarographic dissolved oxygen probe and any other electrodes in use with distilled water blot dry with a lint free tissue and place into the sample 2 Start the measurement and wait for it to stabilize a If the meter is...

Page 7: ...t Data t Data Log t Date Time t Language t Key Press Beep t Alarm Beep t Contrast t Auto Shut Off t User ID t Sample ID t Data Log t Calibration Log t Meter Self Test t Factory Reset t About Meter Read Type Selection 1 In the measurement mode press 2 Press or tohighlightpHChannelorDOChannelandpressf3 select 3 Press or to highlight Mode and Settings and press f3 select 4 Press or to highlight Read ...

Page 8: ...n press f3 info to view the electrode slope between pH buffer points pH Channel only or press f1 back to return to the list of calibrations 8 Press to return to the measurement mode Viewing the Data Log 1 In the measurement mode press 2 Press or to highlight View Log and press f3 select 3 Press or to highlight Data Log and press accept 4 Press or to highlight pH Channel or DO Channel and press f2 ...