26
Warning
Do not overload the trays.
Overloading trays may result in
inadequate sterilization and
unsterile instruments.
Warning
Monitor the sterilization process with
process monitors and biological indi-
cators as described in
Recommended Monitoring (p. 33).
Preparation of Carbon Steel
Instruments
Carbon steel instruments must be carefully prepared prior to
sterilization to prevent rusting. After cleaning according to the
manufacturer's recommendations or the methods presented
in
Instrument Cleaning Prior To Sterilization
, prepare car-
bon steel instruments as follows:
1.
Prepare a 2% solution of sodium nitrate in water
(approximately one table spoon in a quart of
water). Completely immerse the instruments in the
solution and allow them to soak for three minutes.
2.
Remove the instruments and package them for
sterilization or place on trays if sterilizing
unwrapped.
Do not rinse or wipe instruments
prior to sterilization.
Preparing Items for Sterilization
General Preparation Guidelines
1.
Clean items thoroughly before placing them in the
sterilizer. Follow good laboratory or medical proce-
dures for cleaning and decontamination of items.
2.
Place absorbent paper or linen towel in the bottom
of trays before use.
3.
Unwrapped goods should only be processed on
trays. Do not place items on the rack or the bottom
of the sterilizer.
4.
Liquids must not exceed 75% of capacity of flask.
Use self-venting caps.
5.
Open all hinged instruments.
S
TERILIZATION
P
ROCEDURES
Note
Do not place carbon steel instru-
ments directly onto stainless steel
trays. Place an unbleached towel
or paper wrap on the tray and
place the instruments on the towel
or paper.
When wrapping instruments for
sterilization, use a material which
promotes drying.