UltiMate 3000 Series:
SRD-3x00 Solvent Racks
Operating Instructions
Page
17
3 Installation
3.1 Facility Requirements
The installation site must meet the following requirements:
•
The power connection is on the rear panel. Make sure that
♦
Free and unrestricted access to the power connection is ensured at all times.
♦
The connection cable/power cord of the device can be easily reached and disconnected
from the power line at all times. Provide sufficient space behind the device to unplug
the cable.
•
Make sure that the installation site meets the power and environmental specifications
listed in the Technical Information section (
→
page 49).
•
Install the SRD in the laboratory on a stable surface that is free of vibrations. Typically,
the SRD is placed on top of the pump.
•
Make sure that the surface is resistant to solvents.
•
The ambient temperature should be kept as constant as possible.
•
Avoid locations with extreme changes in temperature (such as direct sunlight or drafts)
and high humidity.
•
Allow sufficient clearance behind and to the sides of the SRD for ventilation.
3.2 Unpacking
All electrical and mechanical components of the module are carefully tested before the
instrument is shipped from the factory. After unpacking, inspect the instrument for any
signs of mechanical damage, which might have occurred during transit.
Tips:
Immediately report any shipping damage to both, the incoming carrier and
Thermo Fisher Scientific. Shipping insurance will compensate for the damage
only if reported immediately.
Keep the original shipping container and packing material. They will provide
excellent protection for the module in case of future transit. Shipping the unit
in any other packaging automatically voids the product warranty.
1.
Place the shipping container on the floor and remove the accessories kit and solvent
reservoirs.
2.
Grasp the module by the sides. Slowly and carefully, pull the SRD out of the shipping
container and place it on a stable surface.
Summary of Contents for SRD-3200
Page 2: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Operating Instructions ...
Page 4: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page II Operating Instructions ...
Page 28: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 22 Operating Instructions ...
Page 34: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 28 Operating Instructions ...
Page 44: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 38 Operating Instructions ...
Page 54: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 48 Operating Instructions ...
Page 56: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 50 Operating Instructions ...
Page 60: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 54 Operating Instructions ...
Page 62: ...UltiMate 3000 Series SRD 3x00 Solvent Racks Page 56 Operating Instructions ...