Troubleshooting by User
Error message appear in a red screen.
Figure 19. Example for an error message
Note
You need to pull both release cords in order to unlock both locks.
Note
If problems occur other than those listed in this table, service technician must be contacted.
Failure message
Problem with centrifuge
Possible causes and cures
E-1 - E-97
Restart the centrifuge.
If the message appears again, inform a service
technician.
E-98
The centrifuge cannot be operated.
The run does not start or the centrifuge
brakes and ends the run.
Imbalance detected.
Check the load placed in the rotor.
Check that the bolts of the rotor are well greased.
Restart the centrifuge.
If the message appears again, inform a service
technician.
E- 99
Restart the centrifuge.
If the message appears again, inform a service
technician.