4
Operation
Entering parameters
Thermo Scientific
Sorvall Legend X1R
4-9
3. Press
to confirm the new radius
Setting up a program
1. Enter the program parameters.
2. As described above, please enter braking and acceleration profile, speed/ RCF value, run time,
temperatures, bucket code and bucket radius.
3. Press any of the program store keys for 4 seconds.
Configuring, storing and protecting a program
Programs can be stored and retrieved later by means of the assigned number.
1. Enter the program parameters as described above.
2. Press the program memory key to load any of the programs 1-5.
Press the
key in order to load any of the programs 1-99 by means of the numeric pad.
Enter the program number and confirm with
.
3. You may want to enter a passcode to protect the data.
After four seconds the display shows the message
Program saved
or
Program locked
.
Overwriting a passcode-protected program
If you wish to overwrite a passcode-protected program proceed as follows:
1. Select the program you wish to change.
2. Press the program memory key to load any of the programs 1-5.
Press the
key in order to load any of the programs 1-99 by means of the numeric pad.
Enter the program number and confirm with
.
3. Enter the passcode.
If you enter the wrong password, the following message appears: