Operation
Entering parameters
4-4
Sorvall Legend X1R
Thermo Scientific
4. If available close the rotor with the rotor lid.
5. The display shows:
Entering parameters
Acceleration curve
The Sorvall Legend X1/X1R offers you a total of 9 acceleration and 10 braking curves with which
samples and gradients can be centrifuged.
After the centrifuge is turned on, the last running profile selected is shown.
1. Press the
key once in order to open the acceleration profile selection menu.
The display shows the message
Set acceleration
.
2. Select the desired profile by entering a digit between 1-9 on the numeric pad. Curve number 1 is
the slowest and curve number 9 the fastest.
3. Confirm your entry by pressing
.
Braking curve
1. Press the
kay twice in order to open the braking curve selection menu.
The display shows the message
Set deceleration
:
2. Select the desired curve by entering a digit between 1-9 on the numeric pad. Curve number 1 is the
slowest and curve number 10 the fastest.
3. Confirm your entry by pressing
.
Preselecting Speed / RCF
1. Press the
key.
The display shows the
RPM
or the "RCF"-value depending on the display setting. Press the
key
to toggle between the two modes.
Be sure to check all sealings before starting any aeorsol-tight applications.
See the information in the rotor instruction manual.