background image

Concentrating large quantities of organic solvents, such as hexane, chloroform,
dichloromethane, or acetonitrile from extraction or elution systems are best suit-
ed to the AES1010 and AES2010 SpeedVac Systems.

When processing hazardous samples, install the SCT

120

Chemical Trap on the

exhaust port to capture evaporated solvent. This trap, installed in addition to the
integral refrigerated trap, provides more complete solvent trapping. The SCT120
accepts a variety of disposable cartridges to absorb volatile exhaust. Order the
DTK

120

R Chemical Trap Kit, which includes the trap, disposable cartridge, tubing,

and fittings. These activated-charcoal cartridges trap radioactivity. (The same
technology is commonly used in fume hood holders.)

To install a trap on the exhaust port, remove the muffler assembly from the
VAPOR EXHAUST PORT. Attach a piece of vacuum tubing to this fitting. Attach
the other end to the chemical trap. (Follow enclosed instructions supplied with
the DTK

120

R kit).

The chemical trap requires periodic checking to ensure its continued effective-
ness. Refer to the manual shipped with the trap.

When concentrating samples in ammonium hydroxide, install the ANT100
Ammonia Neutralizing Trap. The disposable ANS121 (Ammonia Neutralizing
Solution) bottles, which screw onto the ANT100, neutralize ammonia gas,
preventing unpleasant odors in the laboratory.

The

CC

120

/DX Deluxe Convenience Cart

is a useful accessory for the envi-

ronmental lab. The SpeedVac

®

resides on the top shelf, while any additional trap-

ping apparatus are placed on the bottom shelf. Addition of the cart produces a
completely mobile concentration system.

The model

DVG

50

Digital Vacuum Gauge

can also be installed. Insertion of

the vacuum gauge tube in the vacuum tubing, on the right hand side leading to
the glass condensation trap, provides vapor pressure reading that can be gener-
ally correlated with the dryness level of the sample. Experience with your specific
test samples will determine the correlation between vacuum levels and sample
dryness.

Test runs are necessary to determine the correct time settings for a given
procedure. To obtain data that is descriptive for both concentration (reducing a
large volume to a small volume) and drying (removing all the solvent), conduct
manual test runs, using the containers, the solvent and the volumes that you will
use for actual samples. Interrupt the run every 15 minutes to measure remaining
sample volume and sample temperature. Continue this test until the test sam-
ples are completely dry. Conduct additional runs at different drying settings.

4.1

SOLVENT HANDLING

4.3

DEVISING PROTOCOLS

4

4.2

ACCESSORIES

Summary of Contents for Savant Speedvac ISS110

Page 1: ...Instruction Manual Thermo Scientific Savant ISS110 Speedvac Concentrator ...

Page 2: ......

Page 3: ... 2008 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries ...

Page 4: ...er Cover 2 3 4 Simple System IntegrityTest 3 Section Application Information Page 4 0 Application Information 3 4 1 Solvent Handling 3 4 4 2 Accessories 4 4 3 Devising Protocols 4 Section Accessories Page 5 0 Accessories 5 Section TroubleShooting Page 6 0 Troubleshooting 5 Troubleshooting Guide 7 8 Section Specifications Page 7 0 Specifications 5 Section Return Page 8 0 To Return an Instrument for...

Page 5: ...es a stable level surface for proper operation Units configured for 115 VAC 60Hz should be plugged into a circuit rated for at least 10 amps Units configured for 230 VAC 50Hz should be plugged into a circuit rated for at least 5 amps IMPORTANT INFORMATION Before operating the unit read section 3 0 Operation and Section 4 0 Application Information to determine your specific application requirements...

Page 6: ...he lid Always balance rotor loads An unbalanced rotor causes vibration that will damage the bearings Load the rotor symmetrically There need not be a tube in each holder but you must be sure to evenly balance the rotor by placing an even number of tubes opposite each other When using a rotor with aluminum tube holders insert all tube holders By using the Drying Rate switch thermal energy can be ap...

Page 7: ...y of the vacuum pump and refrigerated vapor trap 1 Install a clean dry glass condensation flask 2 Allow the refrigerated vapor trap to operate for 45 minutes 3 Open the chamber lid and remove the rotor 4 Fill a plastic beaker that will fit in the chamber with 50 ml of water and place it in the chamber Close the lid 5 Press the drying rate switch to the LOW position 6 Start the system by pressing t...

Page 8: ...121 Ammonia Neutralizing Solution bottles which screw onto the ANT100 neutralize ammonia gas preventing unpleasant odors in the laboratory The CC120 DX Deluxe Convenience Cart is a useful accessory for the envi ronmental lab The SpeedVac resides on the top shelf while any additional trap ping apparatus are placed on the bottom shelf Addition of the cart produces a completely mobile concentration s...

Page 9: ...ion resistant fully adjustable polypropylene shelves and swivel casters DVG50 Digital Vacuum Gauge Ideal for monitoring the vacuum efficiency or sample drying efficiency Insertion into the vacuum line between the rotor cham ber and the refrigerated trap provides vapor pressure readings which can be cor related to sample dryness level Insertion into the vacuum line between the refrig erated trap an...

Page 10: ...ceive 1 an Return material Authorization RMA number and 2 sign the Health and Safety Clearance Form before shipping the equipment back All items returned must be certified to be decontaminated and free of radioactivity When returning equipment that may contain hazardous materials you must pack and label them in accordance with DOT regulations applying to the transportation of hazardous material Yo...

Page 11: ...hermo for assistance 7 SYMPTOM GENERAL No Response System trips circuit breaker or main line fuse Sample does not dry or noticeable change in the drying rate POSSIBLE CAUSES No current to the equipment Power failure at source Short circuit to ground in one of the components Flask cap plugged with ice Internal vacuum pump needs servic ing Vacuum leak Insufficient thermal energy applied to the sampl...

Page 12: ... thaw and refreeze Excessive water ice build up in thermal transfer fluid Flask Cap seated improperly Flask cap worn or fouled SOLUTION EXPLANATION Leave a minimum four inch clearance on all sides Clean the condenser of any accumulated dust or debris CallThermo Scientifice for assistance CallThermo Scientific for assistance Change and clean Glass Condensation Flask daily Leave the main power switc...

Page 13: ...er www thermo com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE È marcato col seguente simbolo Thermo Fischer Scientific ha stipulato contratti con una o diverse società di riciclaggio smaltimento in ognuno degli Stati Membri Europei Questo prodotto verrà smaltito o riciclato tramite ...

Page 14: ... can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE SALES DIVISION Phone 828 658 2711 800 252 7100 FAX 828 645 3368 LABORATORY PARTS and SERVICE Phone 800 438 4851 FAX 828 658 2576 TECHNICAL SUPPORT Phone 800 438 4851 ...

Page 15: ......

Page 16: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 134 3001 00 Rev C ...

Reviews: