background image

The ISS

110

SpeedVac

®

Concentrator is a complete system for solvent

evaporation, sample concentration, and drying using a patented technique that
combines centrifugal force, vacuum, and applied heat. It is intended for use with
aqueous and non-aggressive solvents such as ethanol and methanol. It may also
be used with small quantities of acetonitrile and ammonium hydroxide.

The ISS110 SpeedVac Concentrator combines a sample chamber, a rotor assem-
bly, an oil-free vacuum pump and a refrigerated vapor trap in a small
compact housing. The ISS110 applies vacuum to samples only when the rotor is
spinning, thus preventing bumping and foaming. Where required, the ISS110 can
apply thermal energy to the samples to counteract the cooling effect of
evaporation under vacuum.

Polypropylene fittings and Teflon

®

coated sample chamber and vacuum pump

heads resist corrosion and ensure reliable operation.

UNPACKING.

Open the shipping carton. Compare the contents with the packing

list. Inspect the system and components for any damage from shipment. If
damage is evident, save the shipping carton and immediately notify the carrier or
your distributor.

SITE PREPARATION.

The ISS

110

requires a stable, level surface for proper

operation. Units configured for 115 VAC, 60Hz, should be plugged into a circuit
rated for at least 10 amps. Units configured for 230 VAC, 50Hz, should be
plugged into a circuit rated for, at least, 5 amps.

I

MPORTANT INFORMATION.

Before operating the unit, read section 3.0

Operation and Section 4.0 Application Information to determine your specific
application requirements.

Start-of-day procedure

. At the start of each day, ensure that the refrigerated

trap contains a clean, dry, Glass Condensation Flask (GCF400) and that the sup-
ply of CryoCool

®

Heat Transfer Fluid is adequate.

The CryoCool fluid in the refrigerated trap must be cold before drying samples.

For the best results, maintain electrical power to the system at all times to keep
the refrigerated trap cold and ready for use. When you first turn on the ISS110,
wait 45 minutes for the system to reach operating temperature before processing
samples!

GLASS CONDENSATION FLASK INSTALLATION.

1. Prepare the refrigerated stainless steel trap chamber by adding approximately
800 ml of CryoCool fluid. A line scribed on the wall of the stainless steel trap
indicates the minimum appropriate fluid level (with flask removed from trap).
CryoCool conducts heat away from the Glass Condensation Flask, allowing
vapors to condense on the flask walls.

1.0

INTRODUCTION

2.0

INSTALLATION

3.0

OPERATION

1

Summary of Contents for Savant Speedvac ISS110

Page 1: ...Instruction Manual Thermo Scientific Savant ISS110 Speedvac Concentrator ...

Page 2: ......

Page 3: ... 2008 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries ...

Page 4: ...er Cover 2 3 4 Simple System IntegrityTest 3 Section Application Information Page 4 0 Application Information 3 4 1 Solvent Handling 3 4 4 2 Accessories 4 4 3 Devising Protocols 4 Section Accessories Page 5 0 Accessories 5 Section TroubleShooting Page 6 0 Troubleshooting 5 Troubleshooting Guide 7 8 Section Specifications Page 7 0 Specifications 5 Section Return Page 8 0 To Return an Instrument for...

Page 5: ...es a stable level surface for proper operation Units configured for 115 VAC 60Hz should be plugged into a circuit rated for at least 10 amps Units configured for 230 VAC 50Hz should be plugged into a circuit rated for at least 5 amps IMPORTANT INFORMATION Before operating the unit read section 3 0 Operation and Section 4 0 Application Information to determine your specific application requirements...

Page 6: ...he lid Always balance rotor loads An unbalanced rotor causes vibration that will damage the bearings Load the rotor symmetrically There need not be a tube in each holder but you must be sure to evenly balance the rotor by placing an even number of tubes opposite each other When using a rotor with aluminum tube holders insert all tube holders By using the Drying Rate switch thermal energy can be ap...

Page 7: ...y of the vacuum pump and refrigerated vapor trap 1 Install a clean dry glass condensation flask 2 Allow the refrigerated vapor trap to operate for 45 minutes 3 Open the chamber lid and remove the rotor 4 Fill a plastic beaker that will fit in the chamber with 50 ml of water and place it in the chamber Close the lid 5 Press the drying rate switch to the LOW position 6 Start the system by pressing t...

Page 8: ...121 Ammonia Neutralizing Solution bottles which screw onto the ANT100 neutralize ammonia gas preventing unpleasant odors in the laboratory The CC120 DX Deluxe Convenience Cart is a useful accessory for the envi ronmental lab The SpeedVac resides on the top shelf while any additional trap ping apparatus are placed on the bottom shelf Addition of the cart produces a completely mobile concentration s...

Page 9: ...ion resistant fully adjustable polypropylene shelves and swivel casters DVG50 Digital Vacuum Gauge Ideal for monitoring the vacuum efficiency or sample drying efficiency Insertion into the vacuum line between the rotor cham ber and the refrigerated trap provides vapor pressure readings which can be cor related to sample dryness level Insertion into the vacuum line between the refrig erated trap an...

Page 10: ...ceive 1 an Return material Authorization RMA number and 2 sign the Health and Safety Clearance Form before shipping the equipment back All items returned must be certified to be decontaminated and free of radioactivity When returning equipment that may contain hazardous materials you must pack and label them in accordance with DOT regulations applying to the transportation of hazardous material Yo...

Page 11: ...hermo for assistance 7 SYMPTOM GENERAL No Response System trips circuit breaker or main line fuse Sample does not dry or noticeable change in the drying rate POSSIBLE CAUSES No current to the equipment Power failure at source Short circuit to ground in one of the components Flask cap plugged with ice Internal vacuum pump needs servic ing Vacuum leak Insufficient thermal energy applied to the sampl...

Page 12: ... thaw and refreeze Excessive water ice build up in thermal transfer fluid Flask Cap seated improperly Flask cap worn or fouled SOLUTION EXPLANATION Leave a minimum four inch clearance on all sides Clean the condenser of any accumulated dust or debris CallThermo Scientifice for assistance CallThermo Scientific for assistance Change and clean Glass Condensation Flask daily Leave the main power switc...

Page 13: ...er www thermo com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE È marcato col seguente simbolo Thermo Fischer Scientific ha stipulato contratti con una o diverse società di riciclaggio smaltimento in ognuno degli Stati Membri Europei Questo prodotto verrà smaltito o riciclato tramite ...

Page 14: ... can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE SALES DIVISION Phone 828 658 2711 800 252 7100 FAX 828 645 3368 LABORATORY PARTS and SERVICE Phone 800 438 4851 FAX 828 658 2576 TECHNICAL SUPPORT Phone 800 438 4851 ...