Specifications
Thermo Fisher Scientific
Sarasota FD910, FD950, & FD960 User Guide
B-3
Table B–3.
Process connections
1-inch ASME B16.5 RF
(raised face)
Sarasota FD910 (Class 150, 300, or 600): Stainless steel
(316L/1.4404)
Sarasota FD910 (Class 150, 300, or 600): Duplex (A 182 Gr.F51)
Sarasota FD950 (Class 150, 300, or 600): Hastelloy C276
Sarasota FD960 (Class 150, 300, or 600): Stainless steel
(316L/1.4404)
Sarasota FD960 (Class 150, 300, or 600): Duplex (A 182 Gr.F51)
25-mm BSEN1092 RF
(raised face – type B)
Up to maximum PN100
Other flange types
Consult Thermo Fisher
Table B–4.
Compliance/Certification
Quality assurance
ISO 9001:2000
CE mark
Compliant
Electromagnetic
compatibility
Compliant (EN 61326:1997)
Pressure Equipment
Direct (97/23/EC)
Category III
Low Voltage
Directive
Compliant
Safe area use
As standard
BS EN ISO 15156 /
NACE MR0175
Conformance
Sarasota FD910 and FD950 only
ATEX Conformance,
Intrinsically Safe
(97/9/EC)
F option: EX II 1 G Ex ia IIC T6 Ga (-20°C≤Ta≤+60°C)
H option: EX II 1 G Ex ia IIC T4 Ga (-20°C≤Ta≤+60°C)
ATEX Conformance,
Flameproof (94/9/EC)
EX II 2 G Ex d
b
IIB+H2 T4/T3
Temperature classification of T4 or T3 (for use with maximum
process fluid temperature of +115°C or +180°C respectively).
Canadian Standards
Association (CSA)
Explosion proof Class 1, Groups B, C, & D
Calibration
certification
Calibration traceable to national standards; calibration
certificates supplied as standard. Optional traceable calibration
equipment listing available.
Material traceability
Wetted parts traceability to EN 10204. Type 3.1.
Summary of Contents for Sarasota FD910
Page 2: ......
Page 3: ...Sarasota FD910 FD950 FD960 Liquid Density Meters User Guide P N HB 9135 Revision J...
Page 4: ......
Page 6: ...This page intentionally left blank...
Page 8: ...This page intentionally left blank...
Page 12: ...This page intentionally left blank...
Page 22: ...This page intentionally left blank...
Page 36: ...This page intentionally left blank...
Page 40: ...This page intentionally left blank...