Installation
Electrical Data
F option
At the amplifier terminals 1 & 2: Ui = 30 V, li = 100 mA, Pi = 0.7 W,
Ci=0.01 µF, Li=0
At the PRT terminals W, X, Y, & Z: Ui = 10 V, li = 300 mA per
terminal, Pi = 0.5 W, Ci=0.01µF, Li=0
H option
At the amplifier terminal pairs 1 & 2, 3 & 4, 5 & 6: Ui = 28.5 V,
li = 100 mA, Pi = 0.7 W, Ci=0.01µF, Li=0
(Terminals 5 & 7 are linked internally)
Terminals 7 to 8: Uo = Ui, lo = li, Po = Pi,
Ci=0.01µF, Li=0
Specific Conditions of Use:
1.
The enclosure may be constructed from aluminum alloy and given a protective paint
finish, however, care should be taken to protect it from impact and friction if located in
Zone 0. The painted enclosure constitutes a potential electrostatic charge hazard and
must not be rubbed with a dry cloth or cleaned with solvents.
2.
When the equipment is installed, care must be taken to ensure the maximum
ambient temperature of the electronics housing is not affected by the specified process
temperature of the connected sensor assembly.
Summary of Contents for Sarasota FD900
Page 2: ......
Page 4: ......
Page 6: ...This page intentionally left blank...
Page 8: ...This page intentionally left blank...
Page 12: ...This page intentionally left blank...
Page 36: ......
Page 42: ...This page intentionally left blank...
Page 46: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 58: ...This page intentionally left blank...
Page 68: ...This page intentionally left blank...
Page 88: ...This page intentionally left blank...
Page 89: ......