1-4 RB-XXX-XXXXXX
E RO-20
Löw 08.10.2008
1.4
General Specifications
1.4.1
Ranges
Five linear ranges
•
0–5 mR/h
•
0–50 mR/h
•
0–500 mR/h
•
0–5 R/h
•
0–50 R/h
1.4.2
Meter
•
Scale length: Approx. 3 in. (7.6 cm)
•
2% accuracy.
•
Linear markings: From 0 to 5 in 50 minor increments.
•
Back-lighted.
1.4.3
Response Time
Five seconds, 0 to 90% of final reading.
1.4.4
Linearity
•
Within ± 5% of full scale
•
Battery Dependence
(Within ± 5%)
•
The reading is independent of battery voltage when the battery check indications are in
the green arc. In order to achieve, but linearity in the 50R/h Range, it is recommended
to exchange the 3 ea 12 Volt batteries (indicated by “Battery 2”) once the needle falls
below “4.0”.
1.4.5
Controls
1.4.5.1
External
•
Function Switch with Off, Zero, and Battery checking positions.
•
The Zero Knob is used to set the meter to zero when the Zero position of the Function
Switch is selected, or when the detector is not in a significant radiation field.