WARNING
After running at high current, the anode and cathode plates can become hot. Use caution when separating
the gels and stacks from the plates. When continuously processing multiple samples at ~5A for a maximum of 2 hours,
allow the cassette to cool for 30 minutes or use multiple cassettes to avoid excessive cassette heating.
WARNING
Blotter use outside of the workflows described in this manual may put the operator at risk of dangerous expo-
sure to electrical shock. Do not use this instrument for any purposes or in any configurations not described in this manual.
WARNING
The blotter base unit can contain dangerous electricity and is not designed to opened by the user.
Disconnect all power to the base unit before maintenance by a qualified technician.
WARNING
Do not overfill the cassette with liquid. Excess liquid can overflow into the base unit and possibly cause
electric shock. Follow the appropriate instructions for reagent amounts and empty any remaining liquid in the cassette
upon run completion.
NOTE
A grounded circuit capable of delivering the appropriate current and voltage is required for installation. Electrical requirements can
be located on the rear panel of the blotter base. The blotter electrical system will adjust to the proper voltage for the respective country.
Connect the blotter power cord to the rear left side panel of the device and plug into a grounded power outlet.
Blotter Electrical Parameter
Rating
Supply Voltage (VAC)
100-240
Frequency (Hz)
50/60
Maximum Power Rating (W)
168
Fuse (Power Center)
T3AL, 250V, 3A
ENVIRONMENTALLY FRIENDLY DISPOSAL
According to EU directive 2002/96/EC on electric and electronic equipment and its implementation into national law, all
electric equipments must be separately collected and environmentally friendly recycled. Alternative disposal: If the owner of
the electric equipments does not return it to the manufacturer, he is responsible for proper disposal at a designated collec-
tion point that prepares the device for recycling according to national recycling laws and regulations. This does not include
accessories and tools without electric or electronic components.
For more details, refer to the full blotter user manual available at
www.thermoscientific/pierce
Related Thermo Scientific
™
Products
84731
Pierce 1-Step Transfer Buffer, 1L
84742
Pierce 1-Step Transfer Buffer, 200mL
84783
Western Blotting Filter Paper, 7cm x 8.4cm, 0.83mm thick, 100 mini-sized sheets
84784
Western Blotting Filter Paper, 8cm x 13.5cm, 0.83mm thick, 100 midi-sized sheets
88600
Western Blotting Filter Paper, 8cm x 10.5cm, 0.83mm thick, 100 mini-sized sheets
88018
Nitrocellulose Membrane, 0.45μm, 30cm x 3.5m roll
88518
PVDF Membrane, 0.45μm, 26.5cm x 3.75m roll
2162534 05/13 Printed in China
© 2013 Thermo Fisher Scientific Inc. All rights reserved. Facebook is a registered trademark of Facebook, Inc.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Life Science Research
Africa/Belgium/Europe/Middle East
+32 53 85 71 84
France
0 800 50 82 15
Germany
0228 9125650
Netherlands
076 50 31 880
Switzerland
0800 56 31 40
UK
0800 252 185
Email: [email protected]
thermoscientific.com/pierce
For other regions, visit
thermoscientific.com/piercedistributors
USA
+815-968-0747 or +800-874-3723
Customer Assistance
E-mail:[email protected]
thermoscientific.com/pierce
thermoscientific.com/pierce