24
8.2
Setting Up For A Run (Ready Mode)
KEY
DISPLAY
NOTES
__
1.
NEXT
LOGGING DISABLED
This screen indicates the
logging
status.
To
enable
the
logging
function,
key
ENTER
. Toggling of the
on/off
logging
status
can
be
done
by
keying
ENTER.
2.
ENTER
LOG INTRVL 600s
This message indicates that
TAG#: 4
logging is enabled. Example is
for
10-min
log
period,
selected
through
the
PC
(see
Section
9.0),
and
next
free
tag
is
#4.
3.
NEXT
ALARM: OFF
This screen indicates the
alarm
status.
Keying
ENTER
repeatedly
toggles
through
the 3 alarm modes:
4.
ENTER
ALARM: INSTANT
This enables the alarm based
LEVEL:1.50
mg/m
3
on the real-time concentration.
The
level
(e.g.
1.50
mg/m
3
)
must
be
set
on
the
PC.
5.
ENTER
ALARM: STEL
This enables the alarm based
LEVEL:0.50
mg/m
3
on the 15-min STEL value.
The
level
(e.g.
0.50
mg/m
3
)
must
be
set
on
the
PC.
6.
NEXT
ANALOG OUTPUT:
This screen indicates the analog
DISABLED
signal output status. Keying
ENTER will enable the analog
output. Toggling the analog
output on/off can be done
by
keying
ENTER
:
7.
ENTER
ANALOG OUTPUT:
This enables the analog output.
0
–
0.400
mg/m
3
The concentration range (e.g., 0
–
0.400
mg/m
3
) must be set on
the PC.
8.
NEXT
CAL FACTOR: 1.00
This screen displays the
DIS AVG TIME 10s
calibration factor and the display
averaging
time.
Edit
via
PC