Thermo Scientific Orion VERSA STAR Conductivity/Temperature Module

Module Maintenance

For routine maintenance, dust and wipe the module with a damp cloth. If necessary, warm water or mild water-based 
detergent can be used. Maintenance can be performed on a daily, weekly or monthly basis, as required by operating 
environment. Immediately remove any spilled substance using the proper cleaning procedure for that spill type.

 

Display Information from Module

Example Conductivity Measurement Screen – Single Channel

Date and time

Sample 

temperature

Primary reading

Information on 

last calibration

The information on the display can be customized. 

Enter the setup menu to make changes. 
1. Press 

Setup

.

2. Use the arrow keys to highlight Inst. Settings, 

 

 press 

(f3/select)

 key.

3. Use the arrow key to highlight Display View. 

 

  Press 

(f3/select)

 key.

4. Follow the onscreen prompts to choose what you

 

  would like to see on the measurement screen.

Summary of Contents for Orion Versa Star

Page 1: ...Thermo Scientific Orion VERSA STAR Conductivity Temperature Module Reference Guide...

Page 2: ...riode Minimum Stir Requirement MSR NISS Optimum Results Orion Dual Star Orion Star SAOB SMART AVERAGING SMART STABILITY Star LogR Star Navigator 21 Stat Face Triode are trademarks of Thermo Fisher Sci...

Page 3: ...rement 13 Chapter 7 Methods 15 Chapter 8 Data Storage and Review 17 Data Storage Settings 17 Measurement Read Type 17 Datalog Computer and Printer Export Settings 17 Datalog Review 17 Calibration Revi...

Page 4: ...This page intentionally left blank...

Page 5: ...f this module include Conductivity TDS salinity and resistivity capabilities Selectable reading reference temperatures of 5 10 15 20 or 25 C Compatibility with 2 or 4 cell conductivity cells Linear no...

Page 6: ...This page intentionally left blank...

Page 7: ...sample range and reporting requirements 2 Place conductivity cell in the meter s electrode arm 3 Attach the 8 pin mini DIN connection into the conductivity module 4 Attach the Orion Star stirrer prob...

Page 8: ...tance using the proper cleaning procedure for that spill type Display Information from Module Example Conductivity Measurement Screen Single Channel Date and time Sample temperature Primary reading In...

Page 9: ...2 1 Alarm Limit CalDue Set Point Limit and set point alarm default set to off Cal due alarm set to 12 hours TDS Setup Menu Default values in bold Read Type AutoRead Timed Continuous Cell K Constant I...

Page 10: ...ox is unchecked not selected a single timed measurement will be taken The meter will apply the same saving and exporting data conditions as with the set time interval to record and export data but loc...

Page 11: ...mperature probe to be used in more than one channel for temperature compensated measurements such as when a pH electrode and a 2 cell conductivity cell are in the same measurement sample Temperature i...

Page 12: ...8 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...

Page 13: ...ll user guide Note For automatic calibration the nominal cell constant of the conductivity cell must be entered in the setup menu before the calibration is performed Use fresh Thermo Scientific Orion...

Page 14: ...10 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...

Page 15: ...ents recalibrate if a different conductivity cell is used For temperature calibration the probe needs to be connected to the meter and the calibration solution should have a known stable temperature I...

Page 16: ...12 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...

Page 17: ...t sample 3 If the meter is in AutoRead mode meter default press measure esc 1 key If the meter is in continuous or timed read mode the meter will immediately start taking readings 4 If in AutoRead mod...

Page 18: ...14 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...

Page 19: ...he last calibration data stored for each method Use the channel setup to access and modify methods 1 Press f3 setup key 2 Use the arrow keys to highlight the appropriate channel Press f3 select 3 Use...

Page 20: ...16 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...

Page 21: ...ke your selection 1 From measurement mode press setup 3 2 Press the arrow keys to highlight Inst Settings Press setup 3 3 Use the arrow keys f2 page key and f3 select key to a set the communication pr...

Page 22: ...of the last 10 calibrations for each parameter of the meter The information will match the information recorded for that calibration To review calibration information 1 Press log view 9 key 2 Press ri...

Page 23: ...s in your country and information on Thermo Scientific Orion products that may assist the detection of substances subject to the RoHS Directive are available at www thermoscientific com Error Recommen...

Page 24: ...A Equipment Class Measurement control and laboratory Orion Star A series meters are EMC Class A Conforms with the following directives and standards EN61326 1 2006 Electromagnetic Compatibility EMC Di...

Page 25: ...ation Linear 0 to 10 0 C nLn nLFu EP Compatible Cell Constants 0 001 to 199 9 Calibration Points Up to 6 points Calibration Editing Yes Resistivity Range 2 ohm to 100 meg ohm Resolution 2 ohms cm Rela...

Page 26: ...Set Orion VERSA STAR meter Orion VERSA STAR conductivity temperature module Electrode arm with redesigned holder Universal power adapter VSTAR22 Orion VERSA STAR Conductivity Benchtop Meter Kit Orion...

Page 27: ...th Two pH ISE Modules Kit Orion VERSA STAR meter Two Orion VERSA STAR pH ISE temperature modules 8102BNUWP Orion ROSS Ultra pH electrode 096019 Orion Star stirrer probe 927007MD ATC probe stainless st...

Page 28: ...bottle 810007 ROSS fill solution 5x60 mL Electrode arm with redesigned holder Universal power adapter VSTAR90 Orion VERSA STAR pH ISE Conductivity RDO Dissolved Oxygen Benchtop Multiparameter Meter S...

Page 29: ...rage bottle 011007 conductivity standard 1413 S 5x60 mL Electrode arm with redesigned holder Universal power adapter VSTAR93 Orion VERSA STAR pH ISE Conductivity RDO Dissolved Oxygen Ultimate Benchtop...

Page 30: ...body 927005MD Orion ATC probe epoxy body 013005MD Orion DuraProbe conductivity cell 4 cell K 0 475 013016MD Orion conductivity cell flow through K 0 1 083005MD Orion dissolved oxygen probe polarograp...

Page 31: ...f many aqueous samples Solution 25 C to 50 C Temperature Coefficient C Ultra Pure Water 4 55 Salt NaCl 2 12 5 NaOH 1 72 Dilute Ammonia 1 88 10 HCl 1 32 5 Sulfuric Acid 0 96 98 Sulfuric Acid 2 84 Sugar...

Page 32: ...99 83 56 2 68 59 7 555 822 86 58 3 70 35 7 768 846 88 59 4 72 12 7 983 870 91 61 5 73 91 8 200 894 93 63 6 75 70 8 418 918 96 64 7 77 50 8 638 943 98 66 8 79 32 8 860 968 101 68 9 81 15 9 084 992 103...

Page 33: ...85 1636 170 117 34 130 4 15 10 1665 173 119 35 132 5 15 35 1693 176 121 36 134 6 15 61 1722 179 123 37 136 7 15 86 1751 182 125 38 138 9 16 12 1780 185 127 39 141 0 16 37 1808 188 129 40 143 2 16 63 1...

Page 34: ...30 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module Notes...

Page 35: ...This page intentionally left blank...

Page 36: ...thermo com India Tel 91 22 4157 8800 wai asia thermofisher com Japan Tel 81 045 453 9175 wai asia thermofisher com China Tel 86 21 68654588 wai asia thermofisher com Singapore Tel 65 6778 6876 wai asi...

Reviews: