Thermo Scientific Orion 320, 350 and 370 PerpHecT
®
LogR
®
Meter User Guide
A-1
Appendix A
LogR™ Temperature
Accuracy Tables
Appendix .Table .1 .– . .Two .Point .LogR .Calibration .Average .Temperature .
Compensation .pH .error .at .pH .3 .and .pH .11 .pH .
Electrode
Average Temp.
Error 0 to 25 °C
(°C)
Average Temp.
Error above
20 °C *
(°C)
Average Temp.
Compensation
Error 0 to 25 °C
(pH units)
Average Temp.
Compensation
Error above
20 °C *
(pH units)
8202BN
0.36
0.10
0.004
0.001
8203BN
0.13
0.02
0.002
0.000
8235BN
0.04
0.14
0.000
0.002
8256BN
0.22
0.06
0.003
0.001
8272BN
0.23
0.06
0.003
0.001
9202BN
0.21
0.06
0.003
0.001
9203BN
0.20
0.10
0.002
0.001
9206BN
0.29
0.06
0.003
0.001
9207BN
0.16
0.07
0.002
0.001
9256BN
0.27
0.07
0.003
0.001
9272BN
0.22
0.06
0.003
0.001
* For 20 °C temperature compensation spans .