Maintenance
Replace the Source
The IR source is easily accessible from the bottom of the spectrometer and can be replaced without
having to remove the spectrometer’s cover.
TOOLS AND MATERIALS NEEDED
l
A #1 Phillips head screwdriver
l
Source Replacement Kit
CAUTION
Before replacing the source, turn off the instrument. Replace the source
only with replacement parts supplied by us.
❖
To replace the source
1. Turn off and disconnect the spectrometer.
a. Turn off the instrument and unplug the power cord.
b. Disconnect any cables from the instrument, such as the Ethernet cable or USB devices.
c. If the spectrometer is purged, disconnect the purge lines from the spectrometer. See
d.
CAUTION
Avoid burn hazard.
The source becomes extremely hot while you are using the instrument.
Wait at least 10 minutes after turning the instrument off before you
perform the next step.
62
Nicolet Summit OA Spectrometer User Guide
Thermo Scientific