Thermo Scientific NAPCO 6506 Operator'S Manual Download Page 10

5

PERIODIC MAINTENANCE

1.

If the oven fails to maintain set temperature, it

may indicate a problem with the Temp.
Control, or possibly a heating element failure.
Contact your local service representative.

2.

Sometimes due to shipping, handling and

normal operational wear, oven doors will get
out of alignment. Realignment is
accomplished quickly and easily as follows:

a. Loosen the set screws holding the hinge

bar in place. (Note that the hinge bar moves
easily once the set screws are loosened.)

b. Center the door in the door frame and apply

pressure toward the top of the oven.
Retighten the set screws.

If properly aligned, the door frame should seat

freely against the door gasket without binding
when the oven is placed in the proper position.
In case the latch bar on the right side of the
oven is out of alignment, it can be adjusted in
the same manner.

3.

Gaskets should be checked periodically for

hardening, shrinking, deterioration or other
signs of wear. It is advisable to keep several
gaskets available for use as replacement.

4.

The oven is equipped with silicone door

gaskets, unless specified otherwise. Do not
use silicone door gaskets or vacuum grease
when evaporating solvents in the oven or
when using high analine point oils. BUNA-N
gaskets should be used in applications when
evaporation solvents or when using high
analine point oils.

5.

The interior finish or silicone gaskets may be

damaged by the following:

Diesel oil
JP-4 oil
MIL-0-5606
Hydraulic Fluids
Butyl Acetate
Carbon Tetrachloride
Ethylene Chloride
Gasoline
Kerosene
Zylene
Methyl Chloride
Stoddard Solvent
Toluene
Aromatic Chlorinated solvent

6.

Refer to the following chart as a guide to what

type of door gasket material is best suited for
your needs.

Summary of Contents for NAPCO 6506

Page 1: ...07 5851 Operator s Manual NAPCO Vacuum Oven Thermo Fisher Scientific 401 Millcreek Road Box 649 Marietta Ohio 45750 USA Phone 740 373 4763 Toll Free 800 438 4851 FAX 740 373 4189 ManualP N3177872 Rev...

Page 2: ...TIFICMAKESNOREPRESENTATIONSOR WARRANTIES WITH RESPECT TO THIS MANUAL IN NO EVENT SHALL THERMO BE LIABLE FOR ANY DAMAGES DIRECT OR INCIDENTAL ARISINGOUTOFORRELATEDTOTHEUSEOFTHIS MANUAL For repair infor...

Page 3: ...9 99 4 05 6 06 9 09 11 10 11 12 Update parts list for new kit s manual manufacturelocation updated elec specs to UL report Corrected 3174760 to 3175318 sw Align correct part numbers in parts list Cor...

Page 4: ...tion and Adjustment 2 Specifications 2 Explanation of Controls 3 Temperature Adjustment 4 Vacuum Operation 4 Periodic Maintenance 5 Door Gasket Compatibility Chart 6 Maintenance of Stainless Steel 7 R...

Page 5: ......

Page 6: ...se was carefully packed and thoroughly inspected before leaving our factory Responsibility for safe delivery was assumed by the carrier upon acceptance of the shipment therefore claims for loss or dam...

Page 7: ...Place shelves in chamber with lip up NOTE HEATERS ARE LOCATED ON BOTTOM OF CHAMBER PLACING AN ITEM DIRECTLY ON THE CHAMBER FLOOR MAY CAUSE OVERHEATING OF THE ITEM Install Thermometer A thermometer wel...

Page 8: ...Indicates chamber vacuum in inches of mercury 6 Hose Connections Used for connecting minimum 1 4 vacuum tubing between oven and vacuum pumping systems or other apparatus 7 Valves Used for evacuating c...

Page 9: ...hes of mercury per cubic foot per 24 hours at 29 to 26 inches of mercury Vacuum Sealing Agents High vacuum silicone grease is often necessary to provide a proper seal at tubing connections valve conne...

Page 10: ...on the right side of the oven is out of alignment it can be adjusted in the same manner 3 Gaskets should be checked periodically for hardening shrinking deterioration or other signs of wear It is advi...

Page 11: ...FAIR VERY GOOD Aldehydes FAIR POOR POOR GOOD Alkalis FAIR GOOD GOOD VERY GOOD Amines POOR POOR GOOD Animal Oils EXCELLENT VERY GOOD GOOD Esters Alkyl Phosphate Skydrol POOR POOR GOOD Esters Aryl Phos...

Page 12: ...leaned regularly with mild soapy water and rinsed with distilled water Always thoroughly dry the chamber after cleaning IMPORTANT IF IT IS NECESSARY TO USE THE FOLLOWING CHEMICALS LIMIT THE EXPOSURE T...

Page 13: ...5 Oxalic Acid 2 to 5 in warm water Swab solution on surface allowing it to remain until rust is loosened Immediately flush with copious amounts of clean water 6 A mixture of 20 nitric acid and 1 5 hy...

Page 14: ...inishes Liquid NuSteel Rub with dry cloth Use small amount of cleaner Use in direction of polish lines on No 4 polished finish May scratch No 2 mill and Nos 7 and 8 polished finishes Paste NuSteel DuB...

Page 15: ...ag 57 Rub with dry cloth Use small amount of cleaner Use in direction of polish lines on No 4 polished finish May scratch No 2 mill and Nos 7 and 8 polished finishes Hard Water Spots and Scale Vinegar...

Page 16: ...5 Door Gasket Silicone 3173927 Door Gasket Buna N 3174683 Door Gasket Viton 3167059 DoorAssembly 3166984 DoorLatchAssembly 3167255 Foot Leveler Kit 3174674 SiliconWindowGasket 3173926 Buna NWindow Gas...

Page 17: ...12 WIRING DIAGRAMS 120 VOLT WIRING DIAGRAM...

Page 18: ...13 240 VOLT WIRING DIAGRAM...

Page 19: ...warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo Fisher Scie...

Page 20: ...this warranty Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty pe...

Page 21: ......

Reviews: