Preventative maintenance procedures
Thermo Fisher Scientific
Thermo Scientific™ Multidrop™ Pico 1 Digital Dispenser Operating manual
29
•
The top surface of the bottom cover, below the destination plate
holder.
•
All the outside surfaces of the Multidrop Pico dispenser.
Clean the pogo pins (see
).
For both the pogo pin inspection and cleaning, turn off the Multidrop
Pico dispenser:
∂
Close the Multidrop Pico dispenser software, either by clicking
the X in the upper right corner of the main operating window
or by clicking Exit in the File Menu.
∂
Turn off the Multidrop Pico dispenser.
Remove any unused cassettes from the Multidrop Pico dispenser to
avoid contamination and place them back in the cassette tray.
Place the handle in the down position to lower the pogo pins down.
Carefully, tip the dispenser on its back to be able to see under the
cassette holder.
Inspect the pogo pins through the slot in the cassette holder and ensure
that all pogo pins are not bent and extend fully. Pogo pins that are bent
or not fully extended must be replaced (contact Support).
Tip the dispenser back upright and plug the cables back in.
Pogo pin
inspection