Maintenance
Regular and preventive maintenance
70
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
Figure 5–23.
Removing the reagent filter
l
Place a disposable reagent basin (Cat. no. 9510027) or a 96-well
plate filled with
detergent
, for example, filtered 1% Micro-90
[
International Products Corporation
]
or equivalent, under the
dispensing valves so that the liquid is in contact with the valves
Figure 5–24.
Cleaning the tubings using a reagent basin
l
Press the
EMPTY
button for several minutes.
l
Place another disposable reagent basin or a 96-well plate filled with
distilled and filtered laboratory-grade water
under the dispensing
valves so that the liquid is in contact with the valves
l
Press the
EMPTY
button for several minutes.