8
To Replace the Power Switch:
a) Disconnect mixer from power supply.
b) Peel off the mixing surface disc. Hold the mixing
platform to keep from turning and remove the
platform from the counterweight assembly.
c) Turn mixer upside down and remove the four
screws securing plastic housing. Turn mixer
upright and remove housing.
d) Desolder four wires from switch.
Identify or
mark wires disconnected to insure proper
placement and connection when reinstalling
.
e) Remove defective switch.
f)
Install new switch with the number “3” terminal
on top. Resolder the wires as identified or
marked in Step (d).
g) Reinstall plastic housing and secure with four
screws.
h) While holding the mixing platform to keep it from
turning, secure the platform to the counter-
weight assembly. If the mixing surface is in
good condition, reapply it to the mixing platform.
i)
Reconnect mixer to power supply.
To Replace Rheostat (Speed
Control):
a) Disconnect mixer from power supply.
b) Peel off the mixing surface disc. Hold the mixing
platform to keep from turning and remove the
platform from the counterweight assembly.
c) Turn mixer upside down and remove the four
screws securing plastic housing. Also, remove
two screws securing dial plate.
d) Turn mixer upright and remove the housing.
Maintenance and Servicing
Warning
Disconnect from power supply before
servicing.
Refer servicing to qualified personnel.
Summary of Contents for M37610-33