2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Installing the ESI Probe
Thermo Electron Corporation
LTQ XL Getting Started
39
Installing the ESI
Probe
Install the ESI probe
1. Remove the ESI probe from its storage container. Inspect and clean it if
necessary.
2. Ensure that the probe locking lever on the ion source housing is
unlocked (opened to its widest position). See
3. Insert the ESI probe into the port in the ion source housing, align the
guide pin on the probe body at a minus 45 degree angle from the API
interlock block. See
Figure 13.
Ion Max ion source housing probe locking lever open
Note
If your ESI probe does not already have a sample tube
(fused-silica capillary or metal needle) and safety sleeve attached,
follow the procedure for installing a sample tube and PEEK safety
sleeve that is outlined in
Installing a New Fused-Silica Sample
Tube and PEEK Safety Sleeve
in the
Ion Max API Source Hardware
Manual
.
Probe Locking
Lever (open position)
Sheath Gas Fitting
Probe Port
Grounding Bar
Probe Collar
API Interlock Block
APCI Vaporizer Heater
Cable
Aux Gas Fitting
ESI Interlock Plug
8kV Cable