Optimizing the MS Detector with Your Analyte in APCI/MS Mode
Optimizing the Tune of the MS Detector Automatically in APCI/MS Mode
124
LTQ XL Getting Started
Thermo Electron Corporation
6. On the File/Display toolbar, click the Graph View button to display
the view.
Figure 60.
Tune dialog box, showing the Automatic tuning page
7. Observe the Tune Plus window and the Tune dialog box. While
automatic tuning is in progress, the LTQ XL MS detector displays
various tests in the Spectrum and Graph views in the Tune Plus window
and displays various messages in the Status group box in the Tune dialog
box. Your Tune Plus window should now look similar to the one shown
in
You have now successfully tuned the MS detector in APCI/MS mode for the
compound reserpine (or your analyte of interest). Leave the LC pumps on
(with a flow rate of approximately 400
μ
L/min), and leave the Tune Plus
window open with
APCIhighflow.LTQTune
to complete the next section,