Setting Up the Ion Source for Acquiring Data in APCI/MS/MS Mode
Removing the Ion Max Ion Source Housing
Thermo Electron Corporation
LTQ XL Getting Started
105
Removing the Ion Max
Ion Source Housing
You need to remove the ion source housing to access the ion sweep cone.
Remove the Ion Max ion source housing
1. Remove the drain tube from the ion source housing drain. See
2. Rotate the ion source housing locking levers 90 degrees to release the ion
source housing from the ion source mount assembly.
3. Remove the ion source housing by pulling the housing straight off of the
ion source mount assembly.
4. Place the housing in a safe location for temporary storage.
The Ion Max ion source housing is now properly removed. If you want to
remove the ion sweep cone, go to the next section,
. Otherwise, go to the topic
“Installing the Corona Needle”
Figure 51.
Ion Max ion source housing, detail of components
Note
Disconnect any external liquid lines connected to the ion source
housing before removing the ion source housing.
Ion Source
Housing
Locking Lever(s)
Ion Source
Housing Drain