Setting Up the Ion Source for Acquiring Data in APCI/MS/MS Mode
Removing the Ion Sweep Cone
106
LTQ XL Getting Started
Thermo Electron Corporation
Removing the Ion
Sweep Cone
Use of the ion sweep cone is optional. If you do not need to use the ion
sweep cone it can be removed.
Remove the ion sweep cone
1. Wear a pair of talc-free gloves. Talc-free gloves are required so that the
ion sweep cone is not contaminated.
2. Grasp the outer ridges of the ion sweep cone and pull the cone straight
off of the API cone seal.
3. Store the ion sweep cone in its original shipping container.
The ion sweep cone is now properly removed.
If you need to install the corona needle, go to the next section,
. Otherwise, go to the section,
.
CAUTION
AVOID BURNS.
At operating temperatures, the ion
transfer capillary can severely burn you! The ion transfer tube typically
operates between 200 and 400
°
C.
Always allow the ion sweep cone to
cool to room temperature (for approximately 20 min) before you touch
or remove this component.
Always be careful not to touch the entrance
end of the ion transfer capillary when it is exposed.