Troubleshooting 

WARNING! 

Troubleshooting procedures involve working with high voltages which can cause injury or death. 

Troubleshooting should only be petformed by trained personnel. 

This section 

i s  a 

guide to troubleshooting oven problems. Refer to Table 

for troubleshooting procedures. 

Table 6. 

Problem 

Oven shelves warp and discolor. 

Door gasket comes off. 

System ultimate pressure is high. 

No vacuum in oven. 

Pump stalls. 

Long drying time. 

Probable Causes 

Oven temperature is above 

Gasket is damaged from high temperature 
operation or from resin damage. 

Oven is contaminated with high vapor 
pressure material. 

Pump oil is contaminated. 

Leaks in vacuum piping or ports. 

Vacuum pump seal is worn. 

Discharge valve is worn. 

Internal parts are worn. 

Vacuum vent valve is opened. 

Oven does not pump down properly due to 
misaligned door. 

Power to pump is off. 

Use of extension cord results in insufficient 
amperage. 

Pump is damaged. 

The discharge line is blocked. 

No pump oil or lost viscosity (overheating). 

The piping is too narrow. 

oven is too large. 

Solution 

Do not operate the oven above 

the oven above 

Clean oven to remove resin. 

Clean the oven. Refer to Section 7.1 on page 13. 

Change the oil. Use inlet vapor trap 

gas ballast. 

Check for leaks in vacuum line, oven port, door. 

Replace shaft seal. 

Replace valve. 

Install new vanes. 

Close vent valve. 

Realign the door.  Refer to Section 7.3 on page 

13 

for 

realignment information. 

Check the power switch and outlet. 

Move the oven closer to power source and do not use extension 
cord. 

Return the pump to the factory for repair. 

Change the demister and clear the blockage. 

Change the oil. 

Install piping the same size as the pump inlet. 

Move the pump closer to the oven or use a larger pump. 

Summary of Contents for Lindberg/Blue M VO Series

Page 1: ...f Thermo Fisher Scientific LindbergIBlue M Vacuum Ovens V0914 I V01218 I V01824 Installation and Operating Manual 34598H19 E 11 2011 Part of Thermo Fisher Scientific Therrno S C I E N T I F I C...

Page 2: ...Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any par...

Page 3: ......

Page 4: ...grammingOperation Entering a Program 9 6 3 1 Entering ProgrammingMode 9 6 3 2 Entering Program Parameters 10 6 3 3 RunningaProgram 11 6 3 4 EndingaProgram 11 6 3 5 ChangingaProgram 11 6 4 Auto Tuning...

Page 5: ...electronic control models with suffix P have programmable controls Built in overtemperature protection Fully flexible vacuudpurge release system Vacuum system includes inert gas injection valve fresh...

Page 6: ...y damage can result from processing combustible volatile fluids or materials which emit explosivevapors Voltage 120 VAC 50160Hz Single Phase 240 VAC 50160 H Single Phase 120 VAC 50160 Hz Single Phase...

Page 7: ...I 1 1 1 Power cord Overtemperatureresetbutton WARNING When installing maintaining or removing the refractory insulation the following precautions will minimize airborne dust and fiber Keeppersonnel n...

Page 8: ...he front panel display reads gauge pressure the value is relative to the local atmospheric pressure Use in a location with reduced atmospheric pressure will result in the gauge displaying diminished v...

Page 9: ...nstructions on how to perform the following configuration changes Setting the temperature Setting the OvertemperatureProtection Temperature Changing between celsius and fahrenheit Setting the ramp to...

Page 10: ...and hold SETIENT for 3 seconds to enter the Operating Parameters Menu and display the A1 parameter 10 Usethe Abutton to set the A1 parameter to the desired overtemperature limit in O F 11 Press and re...

Page 11: ...ter 5 Press and release the arrow buttons to show on in the lower display 6 Press SETIENT once to enter the auto tune mode and exit the Operating Parameters menu The controller will cycle three times...

Page 12: ...ler includes an automatic tuning feature and Super control to improve the performance Refer to the UP150 manual for detailed information OFF OFF Pid OFF 32 4 1384 346 1 OFF 0 0 0 ProgramHold Segment A...

Page 13: ...is display mode The SETJENT button will register setpoint value changes until these values are changed again 6 2 3 Exiting Local Mode To exit Single Setpoint or Local Mode and turn off the energy to t...

Page 14: ...to the tM3 display If this value is correct and not changed press and release the SETENT button once to advance to the tM3 display 8 tM3 represents the third time period for the unit to reach the targ...

Page 15: ...time length needed to get to the target setpoint SP1 7 Press and release SETENT button to display other setpoints and segment time lengths Use the arrow buttons to change the temperature setpoints and...

Page 16: ...nstead of the factory setting of C or from O F to C follow these steps These changes will alter the controller input type and associated scale dependant parameters AND ERASE the storedprogram to defau...

Page 17: ...f properly WARNING Disconnect oven from main power A before attemptingany maintenance to oven or its controls Regular maintenance is required to keep the ovenrunning at optimumlevels 1 Clean theovenin...

Page 18: ...t pump down properly due to misaligneddoor Power to pump is off Use of extensioncord results in insufficient amperage Pump is damaged The discharge line is blocked No pump oil or lost viscosity overhe...

Page 19: ...d C Units Door Magnet Catch LevellingFeet AluminumFull Shelf AluminumHalf Shelf 118975 380231 101 38024H01 38025H01 4 38026H01 38028H01 38028H04 Optional Gaskets 38027H01 38023H01 38024H01 38025H01 4...

Page 20: ...KOGAWA UT150 CONTROL NOT TO SCALE ON g05008 I IRE1 Reference V0914 1218 1824 al l a dlrlmTE 5 18 99 T E R A N C E UNLESS OTKRWISE SPKIFIEO 1M VAC EbR A I U 03zn UNLESS OTHERWISE MTEO XX f 032 XXX f 01...

Page 21: ...en finden Sie unter www thermo com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici W...

Page 22: ...sed ___________________________ The above information can be found on the data plate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or spe...

Page 23: ......

Page 24: ...www thermoscientific com...

Reviews: