Diagnostics
Alarm Bad Flags Status
6-6
LevelPRO with
F
OUNDATION
™ fieldbus
Application Guide
Thermo Fisher Scientific
7.
When the maximum current output is shown, enter the value displayed
by the reference meter. Click
Next
.
8.
The software will ask you to confirm the set output and reference meter
reading are the same. Enter
1
to continue to the next step. If you need
to modify the value again, enter
2
and you will return to the previous
screen.
9.
If you entered
1
at the previous screen, the software will show that the
field device is being returned to its original output.
10.
Click
Next
to finish the process.
Alarm Bad Flags
Status
When you select the Alarm Bad Flags Status item from the Diag/Service
menu, any alarms or bad flags (indicators that the gauge has been
incorrectly configured) are displayed. Click
Next
to view each one. An
example is shown below.
Figure 6–8.
Alarm bad flags status example