Diagnostics
Trim Analog Output
The trim analog output function allows you to confirm that the current
output value of the reference meter can be adjusted to match the set value.
Trim Analog
Output
1.
In the NI-FBUS Configurator, go to
AI_TRANSDUCER > Root
Menu > Diag/Service > 2 Trim Analog Output > Trim analog output
.
2.
The first screen appears. Connect the reference meter when prompted.
Click
Next
when you are ready to continue.
Figure 6–3.
Initial Trim Analog Output screen
Thermo Fisher Scientific
LevelPRO with
F
OUNDATION
™ fieldbus
Application Guide
6-3