Safety
In Case of Emergency
4-12
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
In Case of Emergency
shows the plasma door handle:
❖
To shut off the plasma in case of emergency
1. Turn the handle
1
clockwise for 180 degrees.
2. Open the plasma door
2
After 1 second, the plasma switches off.
❖
To shut off the system in case of emergency
1. Switch off the system at the main switch, see
All power to the mass spectrometer is shut off.
2. To separately turn off the computer, use the On/Off switch on the
computer.
Labeled Components: 1=plasma door closed, 2=plasma door open
Figure 4-2.
Plasma door handle at front of iCAP Q MS
1
2