System Description
Gas Supply
3-14
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
Additional gases can be added to the aerosol formed in the spray
chamber. Inside the iCAP Q instrument, tubes are already installed
from the connector to the optional mass flow controller. The connectors
for additional gases are also situated on the left side of the iCAP Q
instrument, see
4
and
5
.
shows the gas outlets at the inlet system.
Plasma and Cooling Gas
details the argon gas requirements:
The argon supply to the iCAP Q system must be stable. A high quality
2-stage regulator at the source is recommended. See
for
suppliers.
Labeled Components: 1=additional gas 2, 2=additional gas 1, 3=nebulizer gas
Figure 3-14.
Gas ports at the inlet system
3
2
1
Table 3-2.
Argon gas requirements for the iCAP Q instrument
Gas requirements
Specification
Purity
99.996% or better
Maximum water content
< 5 ppmv
Supply rate
max. 24 L/min
Pressure 0.6 MPa
(6 bar)