The default configuration settings are as follows:
Step 4 : After checking the settings, terminate the checking mode. [Either
by pressing RUN/STOP Button]:
[Or by pressing the Stirrer Knob]:
Step 5: Terminate configuration mode as described in Configuration
Mode.
The PID parameters for temperature control can be automatically tuned to
take your specific operating circumstances into consideration.
If you select User Mode [Coefficient], the unit will be operated with the
auto tuning value.
The auto-tuning (calibration) procedures are as follows:
Step 1 : Press the CONFIG Button.
Step 2 : Select the auto-tuning mode [CALi] shown
below by turning the Stirrer Knob.
Digital Hotplate 4-23
Thermo Scientific
Section 4
Control Panel
Check Configuration
Settings (LISt) cont.
Auto-tuning (PID
Parameter Calibration)
MODE
Default Setting
MODE
Default Setting
Unit
Cels
Time
ti1
OFST/OFSP
0.0
Limt
100.0
t-H
350.0
t-L
0.0
Coef
Opti
Mmin
1