Appendix 3 SPECIFICATION Page A3-1
Thermo Fisher Scientific
APPENDIX 3
SPECIFICATION
APPENDIX 3
SPECIFICATION
A3.1 Enclosure
Size
Height
300 mm
Width
450
mm
Depth
400
mm
Weight
19 kg
Includes batteries.
Weight will vary slightly depending on cards fitted.
Environment
IP rating
55
Operating temperature*
-10 to +50
o
C
Storage
temperature
-10
to
+70
o
C
Humidity 85%
condensing
*
LCD backlight will not turn on below 0
o
C
A3.2 Power
supply
A3.2.1 AC supply
Voltage
85 to 264 Volts
Frequency
47 to 64 Hz
Power consumption
50 W max
Internal battery
12 V, 13 AH
Internal fuses
20 mm, 3.15AT, input
20
mm,
5AF,
secondary.
A3.2.2 12 Volt DC Supply
Voltage
10.5 to 16 Volts
Fuse
20
mm
5AT
Power consumption for 8 path flowmeter typically 0.4 A @12 V when on continuously, 0.02 A
in sleep mode.
Intermittent modes may be set up by defining on and off times or by an external switch input.
Average consumption is then defined by the on/off ratio.
For example, for an 8 path flowmeter, “On” time 2 minutes, “off” time 13 minutes in a 15
minute cycle, average current = 0.071 A.
A3.2.3 24 Volt DC supply
Voltage
18 to 36 volts
Power consumption
50 W max
Internal battery
12 V, 13 AH
Internal fuses
20 mm, 5AT, input
20
mm,
5AF,
secondary
.
Summary of Contents for HB-S2000