Chapter 3
Carrier Gas Regulation
Gas Control
75
Carrier Gas Regulation
This paragraph describes the automatic Digital Pressure and Flow Control
(DPFC) of the FOCUS GC. You control the gas flow by programming parameters
in the carrier gas menu.
The DPFC module, provides:
•
Flow and pressure control
•
Gas saver function
You enter the gas control setpoints in the
Carrier
menu. Refer to
.
WARNING! Input pressure to DPFC should ensure a minimum pressure 50 kPa higher than the
maximum carrier pressure required by the application.
The maximum input pressure to DPFC is 1000 kPa.
Detector Gas Regulation
The detector gases control module has conventional pneumatic regulators which
and providescontrol for the detector gases.
WARNING! Pressure of the supply gas lines should be set to a minimum of 400 kPa (60 psi) to
ensure the regulation of detector gases.
Maximum input pressure to FOCUS GC (from supply gas lines) is 1000 kPa (145 psi).
The pressure regulation is accessible from outside the instrument. The adjustment
of the detector gases is done in factory prior shipment.
According to the detector in use refer to:
•
•
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...