Chapter 3
Gas Control
Detector Gas Regulation
76
FID Gases
The FID requires hydrogen and air as fuel gas and nitrogen as make-up gas.
The adjustment of the detector gases flow is done in factory prior shipment.
Verify that the corresponding gases flows are correct. If not, adjust the value
turning the relevant regulation screw as shown in Figure 3-1.
Figure 3-1. FID-External Detector Gases Flow Regulation.
The flow is measured at the exit of the detector base body. Enter the FID detector
control menu to turn the flow on/off.
TCD Gases
The TCD detector requires the same gas whether for the measure channel (carrier
and make-up gas, when necessary) and the reference channel (reference gas)
Helium is the recommended carrier gas due to its high thermal conductivity and
chemical inertness.
The adjustment of the detector gases flow is done in factory prior shipment.
Gas
Flow
Hydrogen (fuel gas)
35 mL/min
Air (fuel gas)
350 mL/min
Nitrogen (make-up gas)
30 mL/min
1 turn clockwise of the regulation screw increases the flow of about 2.5 mL/min
1 turn counterclockwise of the regulation screw decreases the flow of about 2.5 mL/min
OPEN
H2
Air
FID - External Detector
Make-up
OPEN
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...