Chapter 3
Gas Control
Carrier Gas Parameters
80
Carrier Gas Parameters
This paragraph explains the electronic programming and control of the GC carrier
gases. The Digital Pressure Flow Control (DPFC) module regulates carrier gas
electronically. You control the gas flow by programming parameters in the carrier
gas menu.
The
CARRIER
menu includes the control parameters for the carrier gas, regardless
of the carrier gas type. Parameters change according to the selected flow mode:
constant flow or constant pressure.
In Main Menu, move the cursor until
Carrier
is selected. Press
ENTER
to open
the
CARRIER
menu.
Flow Mode
Two flow modes are available.
Constant Flow
In constant flow mode, the column flow is kept constant throughout the analysis.
The pressure at the column head will change with the column temperature to
maintain the set flow.
Constant Pressure
In constant pressure mode, the pressure at the column head is kept constant
throughout the analysis. Flow rate will decrease as oven temperature increases.
CARRIER
Flow mode Const flow
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...