Chapter 6
PPKD Injector Menu
Purged Packed Column Injector (PPKD)
119
OPERATING SEQUENCE
Replacing a Septum
Materials required:
•
septum
•
tweezers
WARNING!
T
he injector fittings may be hot. Make sure the injector is at room temperature before
replacing the septum.
1. Remove the septum cap from the injector.
2. Using tweezers, remove the septum from the septum cap.
3. Place a new septum in the septum cap.
CAUTION
To avoid contamination, do not touch the septum with your hands.
4. Gently tighten the septum cap onto the injector assembly until finger-tight.
Do not overtighten the septum cap. The septum will deform and may be
difficult to penetrate with the syringe needle.