To ensure even performance between all channels in a multichannel pipette,
all tip cones have to be changed at the same time, if any of them needs to
be changed. Don’t mix tip cones of different packages, because one bag
contains a matched set of tip cones. Place aligning studs to the same side
when assembling the module.
See picture on page 35
.
There are two versions of battery lids in Finnpipette Novus Pipettes
- Follow the instructions below according to the battery lid version
Opening the Version 1 battery lid
- Use a small fl at screwdriver for opening the battery lid.
- Put the head of the screwdriver into the slot and rotate it to open the lid.
- Disconnect the battery connector and pull out the battery.
Opening the Version 2 battery lid
- Open the battery lid by pressing and pulling the lid as shown in the picture.
- Disconnect the battery connector and pull out the battery.
See page 6
for installing the battery.
Note
Do not reuse version 1 style battery lid
Service
Instructions for
Multichannel
Pipette Tip Cones
Replacing the
battery
1
2
Version 1
Version 2