12
Single Setpoint w/OTP
Models FD1530M, FD1530MCN,
FD1530M-33, FD1535M
Eurotherm 3216 Controller
Operation
The single setpoint model w/OTP furnace controller is
a single setpoint controller, which provides a single
digital display to indicate the current chamber
temperature or setpoint temperature. This temperature
controller features sensor break protection, self-tuning
capability and over temperature protection (OTP) with
an additional OTP relay device.
Basic Operation
When the controller is turned ON it will perform a short
self-test and then display a default page. The
measured value (process value) is found in the upper
display and the setpoint is found in the lower display.
Buttons and Indicators
OP1 (Output 1):
Illuminates when the output is ON
(normally heating).
OP2 (Output 2):
Illuminates when the output is ON
(normally cooling).
OP4 (Output 4):
Illuminates when the AA relay output
is ON (will go on during an alarm situation).
PAGE button:
Allows you to select a new list of
parameters.
SCROLL button:
Allows you to select a parameter
within a list of parameters.
DOWN button:
Allows you to decrease a value.
UP button:
Allows you to increase a value.
PAGE
Button
SCROLL
Button
DOWN
ARROW
Button
UP ARROW
Button
SetPoint (SP)
Display and
Scolling Text
Messages
4 Digit
Process Value
(PV) Display
Unit of Measure
Alarm
Beacon
Output(s)
Status
Beacon
Single Setpoint Models
Note:
If at any time you want to return to the